Common faults and maintenance methods of rotary evaporator/rotary evaporator _ vacuum
Original Title Common Faults and Maintenance wiped film distillation Methods of Rotary Evaporator/Rotary Evaporator Rotary evaporator/rotary evaporator is a kind of common laboratory equipment composed of motor distillation flask heating pot condenser tube and other parts which is mainly used for continuous distillation of volatile solvents under reduced pressure and is often used in chemistry chemical industry biomedicine and other fields The instrument is mainly controlled by electronics and it may break down in the use process which will affect the performance of the product and the results of the experiment so the operator needs to master the solutions to the common faults of the rotary evaporator Rotary evaporator/rotary evaporator The following are the common faults and solutions of some rotary evaporators/rotary evaporators compiled by the editor 1 After the instrument is connected to the power switch The power indicator is not on if the power line is not connected or connected incorrectly just connect the power line; if the power switch or circuit board fails stop using it immediately and contact the manufacturer or professional maintenance personnel 2 The motor does not rotate
If the indicator light of the electric cabinet is on check whether the connection of the plug inside and outside the electric cabinet is loose or broken and then re-insert the plug to connect the broken line; if the indicator light of the electric cabinet or the digital display is not on update the fuse or confirm that the power supply is normal; if the frequency converter is interfered by high frequency and displays “O U” ” remove the fault according to the frequency converter manual 3 The bath is not heated The green indicator light in the temperature controller is on but does not heat and the temperature controller displays “over” or “000” rotary vacuum evaporator which can be solved by replacing the heating ring solid-state relay or relay board temperature controller detecting the probe wiring or updating the probe 4 The vacuum can’t be pumped It is possible that there is solvent in the container and it is limited by the pressure of the saturated ejector so it is necessary to empty the solvent and try again after the bottle is empty; the capacity of the vacuum oil pump is reduced so it is necessary to change the oil (water) and rotovap distillation clean and overhaul the vacuum oil pump; the vacuum hose joint is loose and the vacuum meter leaks so they need to be detected and eliminated along the vacuum pipeline section by section; if there are sundries in the discharge valve and pressure control valve they can be cleaned;